GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
Chapter 10 - Care and Maintenance
10-5
Table 10-2 Customer Care & Maintenance Task Schedule
Service at Indicated Time
Daily
Weekly
Monthly
Annually
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
•
Inspect AC Mains Cable
*
•
*Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Clean Monitor
•
Console Current Leakage Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Peripheral
Current Leakage Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Surface Probe
Current Leakage Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Endocavity Probe Current Leakage
Checks
As prescribed in the Probe Manual.
Transesphongeal (TEE) Probe
Current Leakage Checks
As prescribed in Probe Manual.
Surgical Probe
Current Leakage Checks
As prescribed in Probe Manual.
Measurement Accuracy Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Probe Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Functional Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.