5-42
Section 5-9 - ECG/Respiratory Module
5-9-1-2
Specifications
NOTE:
Respiration function is not recommended for use on paced patients.
Table 5-13 Internal ECG/Respiratory Board Specifications - General
Characteristic
Specification
Ground Isolation
4 kV rms
Table 5-14 Internal ECG/Respiratory Board Specifications - ECG
Characteristic
Specification
Patient Cable
3 leads IEC/AHA codes
Input Range (ac)
10 mV peak-to-peak
Input Range (dc)
Up to ± 850 mV
Bandwidth (-3db)
1.98 — 31.2 Hz (for Monitor use only)
Lead Fault Indicator
Yes
A2D resolution
12 bit, 2’complement
Sample Rate
1000
Table 5-15 Internal ECG/Respiratory Board Specifications - External ECG
Characteristic
Specification
Input Range (ac)
10 V peak-to-peak
Bandwidth (-3db)
0.8 — 65.5 Hz (65.5 Hz in firmware)
Lead Fault Indicator
Yes
A2D resolution
12 bit, 2’complement
Sample Rate
1000
Table 5-16 Internal ECG/Respiratory Board Specifications - Respiration
Characteristic
Specification
Excitation signal characteristics
121 µA RMS ± 5° at 55.5 kHz
Base impedance
100 — 1500 Ohm
Impedance dynamic range
10 Ohm peak-to-peak
Bandwidth (-3db)
0.16 —1.3Hz
Lead Fault Indicator
Yes, Base impedance > 1500 Ohm
A2D resolution
12 bit, 2’complement
Sample Rate
1000