8-24
Section 8-2 - External Component Replacement Procedures
8.) Reconnect the LCD flex cable to the BEP (refer to
).
9.) Refit the Control Panel and Keyboard, as described in the
"Control Panel and Keyboard Installation
.
10.)Close and open the lid a few times to ensure proper closure of the LCD Display cover.
11) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
12.)Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
- only if the Vivid q N system is
currently mounted on the SafeLock cart
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24