GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
Chapter 10 - Care and Maintenance
10-7
Section 10-4
System Maintenance
10-4-1
Preliminary Checks
The preliminary checks take approximately 15 minutes to perform. Refer to the
Vivid q N User Manual
whenever necessary.
Table 10-3 System Preliminary Checks
St
ep
Item
Description
1
Ask & Listen
Ask the customer if they have any problems or questions about the equipment.
2
Paperwork
Note all probes and system options.
3
Power-up
Turn the system power ON and verify that all fans and peripherals turn on. Watch the displays
during power up to verify that no warning or error messages are displayed.
4
Probes
Verify that the system properly recognizes all probes.
5
Displays
Verify proper display on the monitor and touch panel.
6
Presets
Back-up all customer presets on a CD-RW.