Solar 8000M/i patient monitor
4-21
Maintenance and checkout:
Checkout procedure
Keypad/remote control
1. Plug the keypad or remote control into an M-port.
2. Check all functions of the
TRIM KNOB
control and 18 hard keys.
Verify a response at the monitor display.
3. Check that the backlight is on and lights the keys evenly.
4. Activate Boot Code as follows:
Hold down
NBP Go/Stop
and
Zero All
.
Press and release the
TRIM KNOB
control.
Keep holding
NBP Go/Stop
and
Zero All
until the Boot Code
information appears on the display.
5. Select
Tools Menu
.
6. Use password “
mei^
” to open
Service and Diagnostic Tools
Menu
.
7. Select
M-Port Tools
>
Show M-Port #X Serial Incoming Data
.
(Where X = the M-Port number.)
8. If the message, “
No incoming data read
” appears, nothing is
connected to the
M-Port
. Connect the keypad or remote control or
choose the correct
M-Port
.
9. As the columns of numbers scroll, press the keys and
TRIM KNOB
control. Verify that the numbers change as keys are pressed.
10. Exit Boot Code.
Unity Network ID
Unity Network ID connectivity device checkout procedures, see the Unity
Network Interface Device (ID) service manual.
Printer
Press
Graph Go/Stop
and verify that the printer responds correctly.
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......