5-12
Solar 8000M/i patient monitor
2026265-075C
Troubleshooting:
Troubleshooting procedure
Troubleshooting procedure
Many network problems at the central station are traced back to the
patient monitor or Tram-rac housing. Therefore, many of the
troubleshooting suggestions in the “Problems and Solutions” section
include replacing pcb assemblies in the patient monitor or Tram-rac
housing. The “LED Troubleshooting Table” explains the normal
condition of the various LEDs on the pcb assemblies.
Do not perform the procedures listed unless you are a trained field or
biomedical engineer. Refer to
on page 7-7 for circuit board
functions. To aid in disassembly of the unit, refer to
WARNING
Patient data loss. This procedure is intended for use by
service personnel with advanced troubleshooting skills.
The consequences of the following steps may cause loss of
patient data and disruption of the entire Unity Network.
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......