A-22
Solar 8000M/i patient monitor
2026265-075C
Appendix A – Technical specifications:
Solar 8000M/i display
Solar 8000M/i display
The Solar 8000M/i patient monitor is currently available with the
following GE displays:
15-inch, medical-grade, digital flat panel display with touchscreen
15-inch, medical-grade, analog/digital flat panel display with or
without
touchscreen
19-inch, medical-grade, analog/digital flat panel display with or
without
touchscreen
Off-the-shelf (non-medical grade) displays are also compatible (see the
required and recommended specifications in this appendix).
WARNING
Do not connect a monochrome display to the Solar
8000M/i patient monitor. Visual alarm messages may not
appear properly.
WARNING
Use of non-GE displays with the Solar 8000M/i monitor
may result in loss of visual alarms and patient
monitoring. Display functionality must be verified for
proper operation with the Solar 8000M/i patient monitor
in the clinical environment prior to placing into service.
Purchaser’s responsibility
The display purchaser is responsible for meeting Solar 8000M/i display
specifications. GE does not make recommendations regarding specific
display models other than those it offers for sale. For questions regarding
display specifications or compatibility of displays not purchased from
GE, contact the display manufacturer.
The GE warranty only applies to equipment purchased from GE. Service
repairs resulting from failures of equipment not purchased from GE are
billable.
NOTE
An isolation transformer must be used with a computer-grade
display to meet UL and EN specifications.
Medical-grade displays
The Solar 8000M/i patient monitor with a medical-grade display meets
applicable UL and IEC specifications for a medical electrical system. For
this system, an isolation transformer is not required.
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......