Troubleshooting:
Troubleshooting procedure
Tram module patient data
does not display.
1. Verify correct patient or simulator leadwire connections to the Tram module in Tram-rac housing.
2. Verify that the Tram module is completely inserted and ON.
3. Verify that the Tram-rac housing is turned ON.
4. Select the
SOFTWARE REVISION
menu and verify that the
RAC COMM
,
RAC DAS
, or
TRAM
MODULE
software file numbers are there.
If
INCOMPATIBLE SOFTWARE REV
Check for a loose or faulty cable from the patient monitor to the Tram-rac housing.
Cycle power to the Tram-rac housing.
View error logs for TRAM module and patient monitor. Refer to
Replace Tram module with a known good one.
Replace Tram-rac housing with a known good one.
Patient Data Module does
not display patient data.
1. Verify correct patient or simulator leadwire connections to the Patient Data Module.
2. Verify that the Patient Data Module is completely inserted into the docking station and is ON.
3. Check for a loose or faulty cable from the patient monitor to the Patient Data Module.
4. View error logs for the patient monitor. Refer to
5. View error logs for the Patient Data Module. See the Patient Data Module service manual.
6. Replace the Patient Data Module with a known good one.
Patient alarms are not
sounding when activated.
1. If a patient’s waveform and the
DISCHARGED
message both display, consult the operator’s manual
for correct patient admitting instructions. The alarm function remains disabled until the patient is
correctly admitted into the system.
2. Make sure the alarms are enabled. Consult the operator’s manual for correct alarm settings.
3. Verify that the alarm volume is not turned off. Look in
ALARM CONTROL
>
ALARM VOL
:
4. Test if the speaker is functioning, using the
LEARN THE MONITOR
menu option.
Patient alarms are not
sounding at the desired
central station.
The alarm function remains disabled until the patient is correctly admitted into the system.
1. If
DISCHARGED
message displays, consult the operator’s manual for correct patient admitting
instructions.
2. If the monitor was moved from one care unit to another, reprogram the alarm destinations.
Watch for misspelling or spaces in the name.
Make sure that the care unit name programmed into the monitor and central station match exactly.
Typically, the care unit name for a monitor is not shown on the central station display of a patient’s
data. Only the bed number is shown. If the care unit name for a monitor does not match the central
station care unit name, then the monitor identification changes. The monitor identifier shows the
care unit name followed by a vertical slash (
|
) and then the bed number. The solution is to modify
the monitor care unit name to match the central station’s care unit name.
3. Force an alarm from another monitor to test if the speaker is functioning.
Patient data from the Solar
8000M/i does not display at
the central station.
If the central station displays a
NO COMM
message for one
patient, do the following:
Cycle power to the Solar 8000M/i processing unit.
Check for a loose connection between the patient monitor and the Ethernet cable.
View error logs for
TRAM
and the monitor. Refer to
An incorrect LAN address was programmed into the monitor. Refer to
Verify operation of the processor pcb with the “LED Troubleshooting Table.”
Problem
Possible Reason and Solution
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......