Solar 8000M/i patient monitor
4-25
Maintenance and checkout:
Checkout procedure
Cardiac output functions
1. Connect the cardiac output cable adapter to the
Temp/CO
connector
of the acquisition device.
2. Connect the cardiac output adapter to the cardiac output cable
adapter and to the simulator.
3. At the patient monitor, go to
MENUS
⇒
CO
.
4. Set up the monitor:
Set
AUTO MODE:
to
ON
Set
INJECT TEMP:
to
BATH
Set
SIZE:
to
7
Set
INJECT VOL:
to
10CC
Set
COMPUTATIONAL CONSTANT:
to
0.540
5. Select cardiac output menu corresponding to the injectate
temperature. (
CO
)
6. Set the adapter and simulator to
0°
. Verify that the IT value on the
lower left of the display matches the input.
When computing is complete, the
CO
reading on the patient monitor
should read approximately ±5% of simulated value.
7. Repeat this test with the adapter and simulator set to
24°
and the
COMPUTATIONAL CONSTANT:
set to
0.595
.
The computed output should be ±5% of simulated value.
SpO
2
functions (Masimo)
1. Connect the Masimo SpO2 tester cable to the SpO2 sensor adapter
cable.
2. Connect the SpO2 adapter cable to the SpO2 connector on the
acquisition device.
3. Turn the patient monitor on.
4. Verify the following are displayed at the patient monitor: (It might
be necessary to turn the SpO
2
parameter on.)
A waveform with an SpO
2
label.
An SpO
2
% reading between 78- 84%.
A PRR reading between 60 and 62 beats per minute.
5. Disconnect the simulator cable from the acquisition device.
SpO
2
functions (Nellcor OxiMax)
1. Verify that 2 AA alkaline batteries are installed in the NELLCOR
SRC-MAX Pulse Oximetry functional tester.
2. Connect the SRC-MAX to the
SpO
2
connector on the module.
3. On the SRC-MAX, verify that the
IR
and
RED
LED indicators are
both lit.
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......