6-4
Solar 8000M/i patient monitor
2026265-075C
Configuration:
Configuring a patient monitor
Patient monitor type
The
PATIENT-MONITOR TYPE
selection determines the type of
monitor desired, i.e., adult, neonatal or operating room. Different alarms
and parameters are activated for each selection. This menu item is part
of the
SERVICE MODE
menu.
CAUTION
Each time the patient-monitor type is changed, the
ADMIT MENU
function defaults to
STANDARD
configuration. Be aware that some alarms and
parameters may be changed.
NOTE
The keypad or remote control is DIDCA programmed for specific
monitor types. The error message
WARNING: REMOTE
MISMATCHED WITH MONITORING MODE
displays if the
monitor and keypad/remote control do not match.
Access
PATIENT-MONITOR TYPE
option, starting from the
MAIN
menu.
1. Select
MORE MENUS
>
MONITOR SETUP
>
SERVICE MODE
.
2. Enter password using the
TRIM KNOB
control to select the day and
month from monitor screen with leading zeros (e.g., July 4 = 0407).
3. Select
PATIENT-MONITOR TYPE
. Be sure to read the information
in the
ATTENTION
box before changing anything.
4. Rotate
TRIM KNOB
control to select the type of environment the
monitor will be used in.
5. Press
TRIM KNOB
control to exit. Your selection displays at the top
of the screen after the time.
↓
↑
053A
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......