2026265-075C
Solar 8000M/i patient monitor
3-9
Installation:
Front panel connectors and indicators
Unity Network ID connectivity device
The figure below shows the Unity Network ID connectivity device
connected to one of the M-Ports.
NOTE
This connection requires that the Unity Network ID connectivity
device IP address is a
10.X.X.X
address. Refer to the Unity Network
ID connectivity device service manual for more details.
Laser printer
Refer to the instructions provided in Laser Printer Support Kit, pn
2013421-001, for details on interconnection using the interface adapter
and installing a serial card in a laser printer.
WARNING
SHOCK HAZARD. Laser printers are UL 60950/EN
60950 certified equipment, which may not meet the
leakage current requirements of patient care equipment.
This equipment must not be located in the patient
vicinity unless the medical system standard EN
60601-1-1 is followed.
Do not connect a laser printer to a multiple portable
socket outlet (MPSO) supplying patient care equipment.
The use of an MPSO for a system will result in an
enclosure leakage current equal to the sum of all the
individual earth leakage currents of the system if there is
an interruption of the MPSO protective earth conductor.
Ethernet connector
017A
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......