5-26
Solar 8000M/i patient monitor
2026265-075C
Troubleshooting:
Reviewing error/event logs
Useful event data
The event log tracks alarm events and other changes to the system that
can be useful in determining the state of the monitor if an error occurs.
For each event, the event log records the event number, the date, the
time, the event ID, the “bucket,” and an event description. Additionally,
an event description code is logged.
NOTE
“Bucket” data is information regarding the state of the event. For
example, if the monitor’s automatic alarm graphing is turned on or
off, the event bucket will contain the number 0 or 1, which
correspond to off and on.
The table below shows some common events, their event codes, the
corresponding bucket data (if applicable) and other event data (if
applicable). This data can be useful in interpreting the event log.
Event description
Event ID
Bucket data
Other event data
Alarm Graph On/Off Change
800
0 = Alarm graphing turned OFF
1 = Alarm graphing turned ON
Graph Location Change
801
1 = Manual graph location changed
2 = Alarm graph location changed
3 = Print window location changed
4 = 12 lead print location changed
Text describing the graph name is
shown.
Monitor Defaults Change
802
Text describing the default changed
is shown (e.g.,
COLOR FORMAT
).
Alarm Limits Change PPtt
803
Limit is indicated in hex.
Viewing Another Bedside On
804
1 = Split View
2 = Full View
Text describing the viewed bedside
name is shown.
Viewing Another Bedside Off
805
1 = Split View
2 = Full View
Text describing the viewed bedside
name is shown.
Network Connection Change
806
0 = Stops communicating
1 = Starts communicating
TRAM/ECG Module Connection Change
807
0 = Stops communicating
1 = Starts communicating
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......