2026265-075C
Solar 8000M/i patient monitor
4-35
Maintenance and checkout:
Checkout procedure
3. Attach a jumper between pin 1 (Marker Out) and pin 2 (Marker In) of
the
DEFIB SYNC
connector and observe negative spikes in the R-
waves of the displayed ECG waveforms.
4. Remove the jumper.
Marker Out (Frequency)
Signal Pin:
1
Ground Pin:
4
Probe Type:
x10
Time/Division:
0.2S
Volts/Division:
1V
Marker Out (Pulse Width)
Signal Pin:
1
Ground Pin:
4
Probe Type:
x10
Time/Division:
5mS
Volts/Division:
1V
025A
026A
Observe normal R-
waves before the
jumper is installed.
Observe negative spikes in the R-waves while
the jumper is installed. Note that the spikes are
small, and can be difficult to see at times.
040A
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......