2026265-075C
Solar 8000M/i patient monitor
5-3
Troubleshooting:
Terms used
Monitor memory
Monitor memory contains the Ethernet address, internet address, bed
name, care unit name, and standard unit defaults for each monitor.
Protected memory
Protected memory contains a patient’s history and any individualized
changes to the unit defaults.
Power cycle or reboot
To power cycle or reboot the monitor, turn the power OFF at the rear of
the unit, wait 10 seconds, then turn the power ON.
Service mode (main code)
The
SERVICE MODE
menu is found in the monitor
MAIN
menu and is
used for various functions like calibration, video tests, and downloading
monitor interface software. (See details later in this chapter.)
Service menu (boot code)
The Service Menu found in the Boot Code is used when downloading the
Boot Code and main processor code. (See details later in this chapter.)
Warm start (boot code)
A warm start activates the software previously downloaded. The
following steps activates a warm start.
1. Hold down
NBP Go/Stop
and
Zero All
.
2. Press and release the
TRIM KNOB
control.
3. Keep holding
NBP Go/Stop
and
Zero All
until the Boot Code
information appears on the display.
4. Select
Start Patient Monitoring
.
Country selection
This is a Boot Code setting. The choices are
France
,
Germany
, or
Default
which chooses a particular set of GE factory defaults. These
defaults are used when changing the unit type (Adult-ICU, OR, or
Neonatal).
Set language
This is a Boot Code setting. It sets the language for displayed text in the
Main Code. (Do not confuse with
Country Selection
.)
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......