5-2
Solar 8000M/i patient monitor
2026265-075C
Troubleshooting:
Terms used
Terms used
Listed and defined below are terms used in this chapter.
Abort (main code)
This is a Main Code menu selection that may appear on the monitor
during software downloads from the monitor
SERVICE MODE
menu.
ABORT
causes the download to stop by pushing the
TRIM KNOB
control.
Boot loader or boot code
The Boot Loader or Boot Code is used to download software. Entering the
Boot Code does not erase any memory, but downloading new software
will erase protected memory. To activate the Boot Code, perform the
following.
1. Hold down
NBP Go/Stop
and
Zero All
.
2. Press and release the
TRIM KNOB
control.
3. Keep holding
NBP Go/Stop
and
Zero All
until the Boot Code
information appears on the display.
Cold start
A cold start is used only in extreme circumstances. It erases the
protected memory, language, unit name and bed number; automatically
discharges the patient; and restores factory defaults. Ethernet address,
internet address, software feature level, and CRG trend option remain
unchanged.
1. Hold down
NBP Go/Stop
and
Zero All
.
2. Press and release the
TRIM KNOB
control and keep holding
NBP
Go/Stop
and
Zero All
.
3. When the monitor resets and the display goes blank, press and hold
Display On/Off
(in addition to
NBP Go/Stop
and
Zero All
)
until
the message “
performing cold start
” appears on the display.
NOTE
If the Boot Code information appears before the message
“
performing cold start”
appears, begin the cold start again.
Continue (main code)
CONTINUE
is a Main Code menu selection that appears after a
successful software download. It allows the user to continue downloading
other files without resetting the monitor.
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......