2026265-075C
Solar 8000M/i patient monitor
6-15
Configuration:
Advanced user procedures
Set internet address
The internet (IP) address is generated to connect Unity devices for the
exchange of patient data and room-to-room communication. It may be
manually derived from the monitor Ethernet address in this procedure if
the factory-assigned internet address is corrupted.
NOTE
GE highly recommends using a ‘private’ LAN to connect Unity
products. The purpose of the Unity Network is to connect only Unity
devices for the exchange of patient data and room-to-room
communication. Adding non-Unity devices (PCs, laptops, desktops,
etc.) may compromise the ability of the Unity Network to meet its
intended use.
A ‘private’ LAN is not the same as a private IP address.
WARNING
Lost Data. Duplication of an internet (IP) address on a
network will cause lost data. If you change the factory
assigned internet address, you must first record all
other
internet addresses used on your network to avoid
duplication.
Valid and invalid IP addresses
Class A, Class B and Class C IP addresses are valid. Class D and Class E
are invalid.
If using a Class A address, then all other octets (.x) can be any value.
If using a Class B address then all monitors that need to communicate
with each other must have the same first
two
octets. For example:
Class A =
0-127.x.x.x
OK
Class B =
128-191.x.x.x
OK
Class C =
192-233.x.x.x
OK
Class D =
234-239.x.x.x
Invalid
Class E =
240-255.x.x.x
Invalid
130.1.1.2
OK
130.1.3.4
OK
130.1.5.6
OK
130.2.7.8
won’t see other beds
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......