5-24
Solar 8000M/i patient monitor
2026265-075C
Troubleshooting:
Reviewing error/event logs
For some categories of TRAM-NET network errors, two additional lines
are added to the error log entry.
Severity is a measure of how the event/error affected the system. There
are three levels of severity.
Definition of parameters
Name
Description
Network Error
Identifies that a network error occurred.
Channel Number
Identifies the network channel exhibiting the error.
Definition of severity levels
Levels
Description
Continue
The event or error was logged, the task may have or may not
have been finished, but the system was able to go on. Most log
entries will have a severity of
CONTINUE
.
Fatal
The event or error was such that the task is not able to go on.
Recovery was not possible. This always is followed by a
warm
start.
Forced Restart
The system was restarted by a known condition (internet
address change, video test, etc).
Summary of Contents for Solar 8000M/i
Page 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Page 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Page 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Page 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Page 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Page 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Page 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Page 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Page 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Page 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Page 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Page 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Page 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Page 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Page 191: ......