12 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. Lullaby
TM
Warmer
Illustration
Description
Figure 17:
Global Alarm
Global alarm:
Global alarm accompanies all the
alarms except the power failure
alarm for visibility from a distance.
Figure 18:
Check Baby Alarm
Check Baby alarm:
1. Manual Mode:
The Check Baby alarm activates
in the Manual Mode of operation
only when the heater output
has been raised to an output
level
greater
than
25%
continuously for 12 minutes.
The alternating single tone
alarm is activated for this
condition and heater output is
reduced to 25%.
Pressing the Alarm Silence key
silences this alarm and resets
the timer for another 12 minutes
of operation and restores the
heat level to previous setting.
In the Manual Mode, with the
skin probe connected, if the
baby temperature display is
greater than 40°C, the heater
will cut-off and Check Baby
alarm along with the baby
temperature display will flash.
Global
Alarm
Global
Alarm
Check Baby
Alarm
Alarm
silence
key