Symptom
Likely causes
Action
Transformer failure
Replace the transformer
The connector from the
heater tube to the PWA is
damaged.
Replace the PWA
SMPS failure
(for warmer with SMPS and
LED observation lamp to be
added)
Replace SMPS from FRU 2062282-
001
No Mobility
Locked castor brake
Release the castor brake
Damaged castor
Replace the castor
Skin Probe Failure
Defective probe
Replace probe
Improper
Skin
Probe
connection.
Reconnect the Skin Probe
Defective Skin Probe
Replace the Skin Probe
Skin
Probe
is
not
connected.
Connect the Skin Probe
Defective PWA
Replace PWA FRU 2043951-001
NOTE: For Warmer with halogen
lamp and transformer.
Replace PWA FRU 2062277-001
NOTE: For Warmer with SMPS and
LED observation lamp
Defective probe jack on
PWA
Replace PWA FRU 2043951-001
NOTE: For Warmer with halogen
lamp and transformer.
Replace PWA FRU 2062277-001
NOTE: for warmer with SMPS and LED
observation lamp
Heater hood does not
swivel
Hood is not aligned to
center
Align the heater hood to the center
and pull the swivel Knob
Mechanism Jam
Replace with support plate assembly
FRU
Bed base does not tilt
Obstacle
for
tilting
mechanism
Check for any obstacle in the path of
the tilting mechanism
Tilt Knob is loose
Tighten the knob screw.
Bed tilt mechanism failure
Replace the Bed tilt mechanism
Accident
Replace the exteriors
Loose screw
Tighten the screw of cap
Damaged heater hood
exterior
Damaged due to an
accident
Replace the heater hood exteriors
Summary of Contents for Lullaby Warmer
Page 12: ...This page is left blank intentionally ...
Page 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Page 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Page 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...