Lullaby
TM
Warmer 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. 13
Alarms and Indications (Continued)
Illustration
Description
Check Baby alarm (continued):
2. Baby Mode:
The Check Baby alarm activates if
baby temperature deviates from the
set temperature by ± 1°C. The heater
will cut-off if the baby temperature
exceeds the desired value by 1°C. An
alternating single tone alarm is
activated for this condition and the
indicator, which is above the Check
Baby symbol, will illuminate. Once
the baby temperature is within 1°C
of the set point, the alarm will reset
and the heater will turn on once
more.
Pressing the Alarm Silence key
silences this alarm; the visual
indication will stay on until the baby
temperature is within 1°C.
In Baby Mode if the temperature of
the baby stays at less than 30° C for
more than 3 minutes, the Check
Baby alarm activates with both
audio and visual indication. The
heater will also be cut-off.
Pressing the silence key will mute the
audible alarm, switches off the visual
indication
and
restores
the
controlled heating.
Figure 19:
Probe Failure Indicator
Probe Failure/Disconnect Alarm:
The Probe Failure alarm is active during
the Baby Mode of operation.
The alarm turns on when the skin
temperature probe fails, or is disconnected
from the warmer.
The red indicator above the probe failure
symbol illuminates as seen in the figure 19.
In Manual Mode, if the probe fails/ is
removed or disconnected, heat is
maintained with blank baby temperature
display.
In the Baby Mode, during probe failure the
Probe Failure alarm illuminates with a
single tone alternating alarm. The audio
alarm alone can be muted for three
minutes by pressing Alarm Silence key.
Probe
Failure
Indicator
Global
Alarm
Summary of Contents for Lullaby Warmer
Page 12: ...This page is left blank intentionally ...
Page 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Page 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Page 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...