Lullaby
TM
Warmer 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. 31
Symptom
Likely causes
Action
Jammed Push button Key
Replace PWA.
Observation lamp does
not glow
Defective
Observation
lamp
Replace observation lamp.
Lamp switch cable from
the front panel to the PWA
is disconnected.
Check the connection of P1 on PWA
Defective
Observation
lamp socket
Replace cable assembly heater-
observation lamp 2039891-001 in
FRU 2043950-001
NOTE: For Warmer with halogen
lamp and transformer.
Replace
2060529-001
in
FRU
2062276-001 for all 3 configurations
NOTE: For Warmer with SMPS and
LED observation lamp
The lamp switch cable
from the front panel to the
PWA is damaged.
Replace the observation lamp switch
cable harness 2039890-001 in FRU
2043950-001
NOTE: For Warmer with halogen
lamp and transformer.
Replace cable assembly 2060528-
001 in FRU 2062276-001
NOTE: For Warmer with SMPS and
LED observation lamp.
Defective cable assembly
heater-observation lamp
Replace cable assembly heater-
observation lamp 2039891-001 in
FRU 2043950-001
NOTE: For Warmer with halogen
lamp and transformer.
Replace cable assy 2060529-001 in
FRU 2062276-001 for all 230V, 115V,
100V
NOTE: for Warmer with SMPS and
LED observation lamp.
Global Alarm indicator
continues to be lit up.
CPU failure
Replace PWA FRU 2043951-001
NOTE: For Warmer with halogen
lamp and transformer.
Replace PWA FRU 2062277-001
NOTE: for warmer with SMPS and LED
observation lamp
Summary of Contents for Lullaby Warmer
Page 12: ...This page is left blank intentionally ...
Page 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Page 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Page 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...