16 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. Lullaby
TM
Warmer
Figure 24:
Power Failure Indicator
Power Failure Alarm
:
Power Failure alarm turns on when the
external power source fails or is disconnected.
When the Power Failure alarm turns on:
A red indicator located above the Power
Failure alarm symbol is illuminated as
seen in Figure 24.
A continuous audible alarm is activated
at this instance.
Chapter 2
System Setup and Checkout Procedures
2.1
Mechanical Checks
WARNING :-
Before using the Lullaby
TM
Warmer, read this manual. Attempting to use device
without a thorough understanding of its operation may result in patient or user
injury.
WARNING :-
Do not perform the preoperational checkout procedure while the patient occupies
the Lullaby
TM
Warmer.
WARNING:-
Complete the preoperational checkout procedure section of this manual before
putting the Lullaby
TM
Warmer into operation. If the Lullaby
TM
Warmer fails any
portion of the checkout procedure it must be removed from use and repaired.
NOTE
Follow the checkout procedure and the hospital guidelines before admitting a new
patient to the Lullaby
TM
Warmer.
Power
Failure
Alarm
Summary of Contents for Lullaby Warmer
Page 12: ...This page is left blank intentionally ...
Page 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Page 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Page 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...