Route the cable harness assembly with
retaining sleeve into the pillar as shown
in fig 12A. Pull the swivel knob and insert
the three pin connector with harness
into the hood pillar, then route the two
pin connector in the same way, then
insert the retaining sleeve into the hood
pillar with flange oriented as shown in fig
12B.
Route both the harness by pushing
simultaneously. Ensure that harness and
connectors are not damaged.
Note:
While routing the harness if the
connector gets stuck in the slot in the
inside of the pillar, slowly pull it back,
rotate the harness and route it again.
13.
Assemble the retaining sleeve to the rear
bracket assembly as shown Using
M5X10 socket head cap screw with M5
Flat and Spring washers.
14.
After tighten the screw, pull the cable
assembly down from bottom side for the
remaining length.
15.
Close
the
rear
hood
cover.
Mount the hood back on the unit using
the new M8 Loctite Countersunk screws
and M5 socket button head screws
provided.