1.5
Displays
The Lullaby
TM
Warmer has three displays and a bar graph.
Illustration
Description
Figure 12:
Displays
The displays on the control panel
are explained as below:
Baby
temperature
display
indicates
the
temperature
measured
by
the
skin
temperature probe.
Note: The skin probe must be
properly connected to the
warmer and the baby for an
accurate baby temperature
measurement.
Control temperature display
indicates
the
control
temperature the user selects in
the Baby Mode.
The baby temperature range is
from 30.0 – 38.0°C.
The
control
temperature
display is not active in the
Manual Mode.
% Power bar graph indicates
power
level
selected
in
increments of 5%.
APGAR timer display indicates
clocked time in minutes and
seconds up to a maximum of
60 minutes. The APGAR timer
will beep at 1, 5 and 10
minutes, and then every 5
minutes until clocked time of 60
minutes in total is reset to zero.
Baby
Tempera
ture
Control
Temperature
APGAR
Timer
% Power
Bar
graph
Indicator
Global
Alarm
Check
baby
Indicator
Summary of Contents for Lullaby Warmer
Page 12: ...This page is left blank intentionally ...
Page 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Page 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Page 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...