30
Data Log Operations
Data Log is a special application of store mode which enables the calibrator to either
automatically log displays at preset time intervals or to manually log a display on
operation of the
STORE
key. Logged data is written to a user specified file.
To set-up a Data Log file, proceed as follows.
•
Select a task, other than
BASIC
. If using
ADVANCED
, set-up the required output
parameters.
•
Use
SETUP
to select Data Log from the Store Mode Menu (See page 36).
Auto Log (Timer)
Press
STORE
and set-up the Data Log file parameters as shown below. Use
CHANGE
VALUE
(F1) followed by cursor keys to set field values. For Auto Log, set-up
TRIGGER
field
to
PERIODIC
.
Manual Logging
Enter the file details as shown above and select
KEYPRESS
for
TRIGGER
field. Screen
reverts to displayed parameters showing set-up file as shown below:
•
Use a combination of
STORE
and
LOG NOW
(F1) to log events as follows.
Memory Operations
Datalog
Summary of Contents for Druck DPI 615 IS
Page 3: ...K0430 Issue No 2 ii ...
Page 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Page 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Page 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Page 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Page 79: ...K0430 Issue No 2 70 intentionally left blank ...
Page 80: ...71 K0430 Issue No 2 ...
Page 81: ...K0430 Issue No 2 72 ...
Page 82: ...73 K0430 Issue No 2 ...
Page 83: ...K0430 Issue No 2 74 ...
Page 84: ...75 K0430 Issue No 2 ...
Page 86: ...77 K0430 Issue No 2 ...
Page 87: ...K0430 Issue No 2 78 ...
Page 88: ...79 K0430 Issue No 2 ...
Page 89: ...K0430 Issue No 2 80 ...
Page 90: ...81 K0430 Issue No 2 ...
Page 91: ...K0430 Issue No 2 82 ...
Page 92: ...83 K0430 Issue No 2 ...
Page 93: ...K0430 Issue No 2 84 ...
Page 94: ...85 K0430 Issue No 2 ...