53
K0430 Issue No. 2
(9)
Set the loop output current to 5 mA as shown below and check that the output
current, measured by the milliammeter standard is within the limits given in Table
4.
(10)
Repeat (9) for all output values given in Table 4.
Table 4 Electrical Current Output Calibration Tolerances
(11)
Disconnect the calibration/test equipment.
Current Output Calibration, 24mA Range Verification
Tolerances
Applied Current (mA)
Permissible Deviation
0
± 0.05% rdg ± 0.01% FS
5
± 0.05% rdg ± 0.01% FS
10
± 0.05% rdg ± 0.01% FS
15
± 0.05% rdg ± 0.01% FS
20
± 0.05% rdg ± 0.01% FS
24
± 0.05% rdg ± 0.01% FS
20
± 0.05% rdg ± 0.01% FS
15
± 0.05% rdg ± 0.01% FS
10
± 0.05% rdg ± 0.01% FS
5
± 0.05% rdg ± 0.01% FS
0
± 0.05% rdg ± 0.01% FS
VOLTAGE
V
TASK: ADVANCED
SNAPSHOT MODE
VOLTAGE
V
TASK: ADVANCED
SNAPSHOT MODE
VOLTAGE
V
TASK: ADVANCED
SNAPSHOT MODE
Calibration
Summary of Contents for Druck DPI 615 IS
Page 3: ...K0430 Issue No 2 ii ...
Page 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Page 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Page 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Page 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Page 79: ...K0430 Issue No 2 70 intentionally left blank ...
Page 80: ...71 K0430 Issue No 2 ...
Page 81: ...K0430 Issue No 2 72 ...
Page 82: ...73 K0430 Issue No 2 ...
Page 83: ...K0430 Issue No 2 74 ...
Page 84: ...75 K0430 Issue No 2 ...
Page 86: ...77 K0430 Issue No 2 ...
Page 87: ...K0430 Issue No 2 78 ...
Page 88: ...79 K0430 Issue No 2 ...
Page 89: ...K0430 Issue No 2 80 ...
Page 90: ...81 K0430 Issue No 2 ...
Page 91: ...K0430 Issue No 2 82 ...
Page 92: ...83 K0430 Issue No 2 ...
Page 93: ...K0430 Issue No 2 84 ...
Page 94: ...85 K0430 Issue No 2 ...