K0430 Issue No. 2
mA Step
To select one of the electrical output programs, press the
OUTPUT
key and proceed as
follows:
On selection of (e.g.) Linear, the output display window changes to show the selected
program of output currents:
•
Connect an external power source as shown on page 24.
•
Press
RUN
(F1) to run program. A flashing status display
CHECK LOOP
indicates a
fault in the external loop i.e. supply fault or open circuit.
Note:
The dwell time at each step is approximately 10 seconds.
•
Press
STOP
(F1) when running to stop at any point. Press
RUN
(F1) to resume.