Current Input Range (55 mA)
Use the following procedure for calibrating the current input range.
(1)
Switch the instrument on, enter
CALIBRATION
mode and select
INT RANGES
from
the
CALIBRATION
menu as shown on page 43.
(2)
Select
CURRENT IN
range for calibration:
(3)
Open circuit the
mA in
terminals and enter the zero point:
(4)
Connect the mA in terminals of the instrument to a variable current source and
connect a digital milliammeter in series with the supply:
55mA
Vin
mA in
max 30V
max
+
-
A
Calibration