55
K0430 Issue No. 2
Calibrate External Sensors
Use the following procedure for calibrating external pressure sensors.
(1)
Connect the required external transducer to the EXT TRANSDUCER socket located
on the rear of the instrument.
(2)
Allow the instrument’s temperature and the temperature of the external
transducer to stabilize in the calibration environment for a minimum of 1 hour.
(3)
Switch the instrument on, enter
CALIBRATION
mode and select
EXT SENSORS
from the
CALIBRATION
menu:
(4)
Select the transducer to be calibrated from the transducer menu using the
cursor keys and press ENTER:
If the sensor to be calibrated is not in the directory, or no sensors are listed in the
directory, press the
F1
key (ADD NEW SENSOR). This will place the sensor in the
directory, allowing it to then be selected.
(5)
Connect the pressure standard to the inlet of the external transducer, apply the
zero point pressure and store the zero point:
INT RANGES
CHANGE PIN
EXT SENSORS
Calibration
Summary of Contents for Druck DPI 615 IS
Page 3: ...K0430 Issue No 2 ii ...
Page 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Page 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Page 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Page 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Page 79: ...K0430 Issue No 2 70 intentionally left blank ...
Page 80: ...71 K0430 Issue No 2 ...
Page 81: ...K0430 Issue No 2 72 ...
Page 82: ...73 K0430 Issue No 2 ...
Page 83: ...K0430 Issue No 2 74 ...
Page 84: ...75 K0430 Issue No 2 ...
Page 86: ...77 K0430 Issue No 2 ...
Page 87: ...K0430 Issue No 2 78 ...
Page 88: ...79 K0430 Issue No 2 ...
Page 89: ...K0430 Issue No 2 80 ...
Page 90: ...81 K0430 Issue No 2 ...
Page 91: ...K0430 Issue No 2 82 ...
Page 92: ...83 K0430 Issue No 2 ...
Page 93: ...K0430 Issue No 2 84 ...
Page 94: ...85 K0430 Issue No 2 ...