15
K0430 Issue No. 2
Pressure to Display (P-DISPLAY) Task
P-Display is a special application of Data Log. To use this mode, select Data Log from the
Store Mode menu as detailed on page 36. Connect the UUT to the instrument as shown
below and, if required, turn on and set-up Cal Mode (see page 11).
A - External pressure source
B - Pressure regulator
C - Dial gauge under test
•
Press
TASK
and select
P-DISPLAY
. I
f required, use
OUTPUT
key to change
pressure units.
•
Set-up a data log file as detailed on page 30.
Note: TRIGGER
field, automatically set to
KEYPRESS
, cannot be changed.
•
Apply a series of test pressures to the device under test. Enter displayed
reading at each pressure and log each point:
•
After logging final test point, terminate as follows:
A
B
PRESSURE INT
bar
F2
F1
DISPLAY
CHANGE
VALUE
TASK : P-DISPLAY
SNAPSHOT MODE
max 30V
DPI 615 IS
C
Taking Measurements
Summary of Contents for Druck DPI 615 IS
Page 3: ...K0430 Issue No 2 ii ...
Page 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Page 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Page 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Page 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Page 79: ...K0430 Issue No 2 70 intentionally left blank ...
Page 80: ...71 K0430 Issue No 2 ...
Page 81: ...K0430 Issue No 2 72 ...
Page 82: ...73 K0430 Issue No 2 ...
Page 83: ...K0430 Issue No 2 74 ...
Page 84: ...75 K0430 Issue No 2 ...
Page 86: ...77 K0430 Issue No 2 ...
Page 87: ...K0430 Issue No 2 78 ...
Page 88: ...79 K0430 Issue No 2 ...
Page 89: ...K0430 Issue No 2 80 ...
Page 90: ...81 K0430 Issue No 2 ...
Page 91: ...K0430 Issue No 2 82 ...
Page 92: ...83 K0430 Issue No 2 ...
Page 93: ...K0430 Issue No 2 84 ...
Page 94: ...85 K0430 Issue No 2 ...