K0430 Issue No. 2
iv
Electrical Parameters
Maximum Output Parameters at the External Measuremen
t
Connectors:
The output parameters at sockets SK1, SK2 and SK3 do not exceed the values specified
in Clause 5.4, Simple Apparatus, of EN 50020.
Maximum Safe Input Parameters:
Sockets SK1, SK2, SK3 and SK6
U
i
= 30V
l
i
= 100 mA
P
i
= 1.0W
Installation
WARNING: Do not use tools on the pressure calibrator that might cause
incendive sparks - this can cause an explosion.
• Provide additional protection for equipment that may be damaged in service
.
• Installation should be carried out by qualified plant installation technicians in
compliance with the latest issue of EN 60079-14.
Declaration Requirements
The DPI 610 IS and DPI 615 IS pressure calibrators are designed and manufactured to
meet the essential health and safety requirements not covered by the EC Type
Exam
i
nation Certificate BAS02ATEX1174X when installed as detailed above. The
intrinsically safe pressure calibrators are designed and manufactured to protect against
other hazards as defined in paragraph 1
.
2
.
7 of Annex 11 of the ATEX Directive 94/9/EC.
lin (SK1)
Vin (SK2)
SwitchIn (SK3)
U
o
= 1.1V d.c.
I
o
= 0
.
16 mA d.c.
P
o
= 0.15 mW
C
i
= 0.05 µF
L
i
= 0
U
o
= 1.1V d.c.
I
o
= 0
.
11 µA d.c.
P
o
= 0.03 µW
C
i
= 0
L
i
= 0
U
o
= 1.1V d.c.
I
o
= 12 mA d.c.
P
o
= 11 mW
C
i
= 0.05 µF
L
i
= 0
lout (SK6)
RS232
External Transducer
U
o
= 7.9V d
.
c.
C
i
= 0
L
i
= 0.1 mH
U
o
= 7.6V d
.
c.
I
o
= 82 mA d.c.
P
o
= 162 mW
C
i
= 0
L
i
= 0
U
m
= 250V
U
o
= 7.6V d
.
c.
I
o
= 155 mA d.c.
P
o
= 0.43 W
C
i
= 0.15 µF
L
i
= 0.9 mH
C
o
= 8.6 µF
L
o
= 0.3 mH
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for Druck DPI 615 IS
Page 3: ...K0430 Issue No 2 ii ...
Page 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Page 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Page 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Page 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Page 79: ...K0430 Issue No 2 70 intentionally left blank ...
Page 80: ...71 K0430 Issue No 2 ...
Page 81: ...K0430 Issue No 2 72 ...
Page 82: ...73 K0430 Issue No 2 ...
Page 83: ...K0430 Issue No 2 74 ...
Page 84: ...75 K0430 Issue No 2 ...
Page 86: ...77 K0430 Issue No 2 ...
Page 87: ...K0430 Issue No 2 78 ...
Page 88: ...79 K0430 Issue No 2 ...
Page 89: ...K0430 Issue No 2 80 ...
Page 90: ...81 K0430 Issue No 2 ...
Page 91: ...K0430 Issue No 2 82 ...
Page 92: ...83 K0430 Issue No 2 ...
Page 93: ...K0430 Issue No 2 84 ...
Page 94: ...85 K0430 Issue No 2 ...