K0430 Issue No. 2
42
Test Equipment
The calibration procedures require the following test equipment.
*ppm = parts per million
Table 1 Internal/External Pressure Transducer Verification
Test Equipment and Instrument Parameter/Range
Calibration
Equipment
Accuracy
Calibration
Uncertainty
Digital Voltmeter - 5V input
±30 *ppm ±1 digit
±10 *ppm ±5µV
Digital Voltmeter - 30V input
±45 *ppm ±1 digit
±11 *ppm ± 110 µV
Digital milliammeter - 55mA input
±150 *ppm ±4 digit
±100 *ppm ± 1nA
Digital milliammeter - 24mA output
±150 *ppm ±4 digit
±160 *ppm ± 1nA
Precision thermometer - ambient temperature
±0.2°C
± 0.1°C ±1 digit
Deadweight tester - pressure internal/external
Class A deadweight
<0.025%
Nominal applied value as a %
of ±full-scale
Permissible deviation
0% (10% for absolute ranges)
± 0.025% FS
20%
± 0.025% FS
40%
± 0.025% FS
60%
± 0.025% FS
80%
± 0.025% FS
100%
± 0.025% FS
80%
± 0.025% FS
60%
± 0.025% FS
40%
± 0.025% FS
20%
± 0.025% FS
0% (10% for absolute ranges)
± 0.025% FS
Calibration
Summary of Contents for Druck DPI 615 IS
Page 3: ...K0430 Issue No 2 ii ...
Page 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Page 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Page 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Page 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Page 79: ...K0430 Issue No 2 70 intentionally left blank ...
Page 80: ...71 K0430 Issue No 2 ...
Page 81: ...K0430 Issue No 2 72 ...
Page 82: ...73 K0430 Issue No 2 ...
Page 83: ...K0430 Issue No 2 74 ...
Page 84: ...75 K0430 Issue No 2 ...
Page 86: ...77 K0430 Issue No 2 ...
Page 87: ...K0430 Issue No 2 78 ...
Page 88: ...79 K0430 Issue No 2 ...
Page 89: ...K0430 Issue No 2 80 ...
Page 90: ...81 K0430 Issue No 2 ...
Page 91: ...K0430 Issue No 2 82 ...
Page 92: ...83 K0430 Issue No 2 ...
Page 93: ...K0430 Issue No 2 84 ...
Page 94: ...85 K0430 Issue No 2 ...