K0430 Issue No. 2
50
(5)
Set the input current to 55 ±0.1mA and enter the measured full-scale input
current:
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to
quit the
CALIBRATION
and
SETUP
modes.
(7)
Verify the instrument calibration by applying the test currents given in Table 3 to
the milliammeter.
(8)
Disconnect the calibration/test equipment.
Table 3 Electrical Current Input Calibration Tolerances
Current Calibration, 55mA Range Verification Tolerances
Applied Current (mA)
Permissible Deviation
-55
± 0.05% rdg ± 0.004% FS
-45
± 0.05% rdg ± 0.004% FS
-35
± 0.05% rdg ± 0.004% FS
-25
± 0.05% rdg ± 0.004% FS
-15
± 0.05% rdg ± 0.004% FS
-5
± 0.05% rdg ± 0.004% FS
0
± 0.05% rdg ± 0.004% FS
5
± 0.05% rdg ± 0.004% FS
15
± 0.05% rdg ± 0.004% FS
25
± 0.05% rdg ± 0.004% FS
35
± 0.05% rdg ± 0.004% FS
45
± 0.05% rdg ± 0.004% FS
55
± 0.05% rdg ± 0.004% FS
Calibration
Summary of Contents for Druck DPI 615 IS
Page 3: ...K0430 Issue No 2 ii ...
Page 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Page 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Page 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Page 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Page 79: ...K0430 Issue No 2 70 intentionally left blank ...
Page 80: ...71 K0430 Issue No 2 ...
Page 81: ...K0430 Issue No 2 72 ...
Page 82: ...73 K0430 Issue No 2 ...
Page 83: ...K0430 Issue No 2 74 ...
Page 84: ...75 K0430 Issue No 2 ...
Page 86: ...77 K0430 Issue No 2 ...
Page 87: ...K0430 Issue No 2 78 ...
Page 88: ...79 K0430 Issue No 2 ...
Page 89: ...K0430 Issue No 2 80 ...
Page 90: ...81 K0430 Issue No 2 ...
Page 91: ...K0430 Issue No 2 82 ...
Page 92: ...83 K0430 Issue No 2 ...
Page 93: ...K0430 Issue No 2 84 ...
Page 94: ...85 K0430 Issue No 2 ...