K0430 Issue No. 2
44
Internal Pressure Range
Use the following procedure for calibrating the internal pressure range.
Note:
If calibrating the hydraulic calibrator version, the calibrator must first be primed as
detailed on pages 57 to 64.
(1)
Connect the outlet port of the instrument to a pressure standard.
(2)
Allow the instrument’s temperature to stabilize for a minimum of 1 hour.
(3)
Switch the instrument on, enter
CALIBRATION
mode and select
INT RANGES
from
the
CALIBRATION
menu and then
PRESSURE INT
.
(4)
Apply zero pressure and store the zero point:
(5)
Close the vent valve, apply full-scale pressure and store the full-scale point:
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to
quit
CALIBRATION
and
SETUP
modes.
(7)
Check calibration by applying test pressures in Table 1, page 42.
VOLTAGE IN
Calibration
Summary of Contents for Druck DPI 615 IS
Page 3: ...K0430 Issue No 2 ii ...
Page 66: ...57 K0430 Issue No 2 Hydraulic Calibrator Versions 3 ...
Page 67: ...K0430 Issue No 2 58 This page intentionally left blank ...
Page 73: ...K0430 Issue No 2 64 This page intentionally left blank ...
Page 74: ...65 K0430 Issue No 2 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 78: ...69 K0430 Issue No 2 Appendix 2 ATEX Certificate of Conformity ...
Page 79: ...K0430 Issue No 2 70 intentionally left blank ...
Page 80: ...71 K0430 Issue No 2 ...
Page 81: ...K0430 Issue No 2 72 ...
Page 82: ...73 K0430 Issue No 2 ...
Page 83: ...K0430 Issue No 2 74 ...
Page 84: ...75 K0430 Issue No 2 ...
Page 86: ...77 K0430 Issue No 2 ...
Page 87: ...K0430 Issue No 2 78 ...
Page 88: ...79 K0430 Issue No 2 ...
Page 89: ...K0430 Issue No 2 80 ...
Page 90: ...81 K0430 Issue No 2 ...
Page 91: ...K0430 Issue No 2 82 ...
Page 92: ...83 K0430 Issue No 2 ...
Page 93: ...K0430 Issue No 2 84 ...
Page 94: ...85 K0430 Issue No 2 ...