K0430 Issue No. 2
46
(5)
Set the variable supply voltage to 5V ± 0.1V and enter the measured full-scale
voltage applied:
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key once to return
to the calibration menu or four times to quit the
CALIBRATION
and
SETUP
modes.
(7)
Verify the instrument calibration by applying the test voltages given in Table 2
(page 48), to the voltmeter (after both voltage ranges have been calibrated).
(8)
Disconnect the calibration/test equipment.
Calibration
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com