Chapter 3. Programming
50
PanaFlow™ LC User’s Manual
Figure 59: Setting up Frequency Output
3.5.3c Understanding the Error Handling Option
The Table 10 shows the response to each of the Frequency Output Error Handling options. For a multi-path
meter, the
[Path Error Handling]
set to ON (see Figure 70) changes the Output response. See Table 11 for
Analog Output response with
[Path Error Handling]
set to ON.
Note:
Table 11 assumes Composite Actual Volumetric is chosen as Measurement for Frequency Output and
the Path configuration is 3 paths.
Table 10: Frequency Output Error Handling options
Option
Output Response
Low
Forces Output to 0Hz on error
High
Forces Output to 10000Hz on error
Hold
Holds the last “good” Hz reading
Other
Enables the user to enter a value between 0Hz and 12000Hz to be
output during an error
Digital Output (Std)
Off
Alarm
Measurement Type
Channel 1
Channel 2
Channel 3
Composite
Measurement Type/Comp
Pulse
Frequency
Base Value
Full Value
Full Frequency
Measurement Type
Velocity
Volumetric (Act)
Soundspeed
Volumetric (Std)
Avg Vol Flow Rate
Mass Flow
Mass Flow
Base Value
UNIT: m/s
0.000
Test Frequency
Base Value
Base Value
0.000 m/s
Frequency
Full Value
Measurement Type
Full Frequency
Frequency
Test Frequency
Full Value
Error Handling
Full Frequency
Test Frequency
0 Hz
Error Handling Value
Frequency
Full Value
Full Frequency
Error Handling
Low
High
Hold
Other
Measurement Type/Comp
Measurement Type/CH #1
Reynolds#
SNR Up
Gain
Std Deviation Gain
Measurement Type/CH #1
SNR Dn
Measurement Type
Composite
Channel 1
Channel 2
Channel 3
Velocity
Volumetric (Act)
Mass Flow
Soundspeed
Test Frequency
0
UNIT: Hz
Test Frequency
12000 Hz
Value
UNIT: Hz
12000
Frequency
Base Value
Error Handling
Value
9
2
9
2
9
2
9
2
9
2
9
2
9
2
or
9
2
9
or
9
2
9
or
9
2
9
or
9
2
9
or
9
2
9
Input/Output
Digital Output (Std)
Analog Output (Std)
Option Comm (Slot 1)
Option I/O (Slot 2)
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
Operator Access or Higher
Summary of Contents for Baker Hughes PanaFlow LC XMT1000
Page 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii PanaFlow LC User s Manual ...
Page 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Page 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Page 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Page 189: ......