Chapter 4. Error Codes and Troubleshooting
86
PanaFlow™ LC User’s Manual
A6:(S2:3)Ch Not
Calibrated
Error occurs when Analog
Input/RTD(S2:3) are not
calibrated
Calibrate the Analog Input/RTD input. If error
persists after calibration, contact BHGE factory
A7:(S2:4)Ch Not
Calibrated
Error occurs when Analog
Input/RTD (S2:4) are not
calibrated
Calibrate the Analog Input/RTD input. If error
persists after calibration, contact BHGE factory
A10:(S2:3)Input
NotConnect!
Analog Input: Error occurs
when (4-20mA) input is not
connected at Channel (S2:3).
RTD Input: Error occurs when
RTD input is not connected or
temp greater than 390 deg C
at Channel (S2:3)
Check connectivity for Analog Input/RTD input
and RTD temperature. Try power cycling the
meter. If error persists after power cycle, contact
BHGE factory
A11:(S2:4)Input
NotConnect!
Analog Input: Error occurs
when (4-20mA) input is not
connected at Channel (S2:4).
RTD Input: Error occurs when
RTD input is not connected or
temp greater than 390 deg C
at Channel (S2:4)
Check connectivity for Analog Input/RTD input
and RTD temperature. Try power cycling the
meter. If error persists after power cycle, contact
BHGE factory
A12:(S2:3)Ch
OverRange Err!
Exceeds input values. For
analog input (S2:3) greater
than 21mA
Ensure analog input current less than 21mA. Try
power cycling the meter. If error persists after
power cycle, contact BHGE factory
A13:(S2:4)Ch
OverRange Err!
Analog input(S2:4) greater
than 21mA
Ensure analog input current less than 21mA. Try
power cycling the meter. If error persists after
power cycle, contact BHGE factory
A24:Aout(S2:1)OutOfRa
nge!
When output from analog
output(S2:1) exceeds 21 mA
or less than 3.6 mA
Check the flow velocity. If velocity is within
limits and error still persists, contact BHGE
factory
A25:Aout(S2:2)OutOfRa
nge!
When output from analog
output(S2:2) exceeds 21 mA
or less than 3.6 mA
Check the flow velocity. If velocity is within
limits and error still persists, contact BHGE
factory
A31:(S2:3)Ch
UnderRange!
Lesser input values. For
analog input(S2:3) between
3.6 mA to 0.25mA
Check input analog current is between 3.6 mA to
21mA. If error persists, contact BHGE factory
A32:(S2:4)Ch
UnderRange!
Lesser input values. For
analog input(S2:4) between
3.6 mA to 0.25mA.
Check input analog current is between 3.6 mA to
21mA. If error persists, contact BHGE factory
Error Code
Error Message
Description
Summary of Contents for Baker Hughes PanaFlow LC XMT1000
Page 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii PanaFlow LC User s Manual ...
Page 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Page 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Page 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Page 189: ......