Chapter 4. Error Codes and Troubleshooting
4.11 Diagnostics Data
To determine the health of the meter, PanaFlow™ LC has built-in diagnostic parameters. Please refer to
Table 28 below for diagnosing any problems with the system. If the meter shows errors and the diagnostics
data indicate issues, fill in the User/Service record appendix before contacting BHGE factory.
Table 28: Diagnostic Parameter Description and Health Indicators
Parameter
Description
Good Bad
Sound Speed
Measured speed of
sound of the fluid
•
Under ideal conditions
sound speed should be
within 5 ft/s (1.5 m/s)
between channels.
•
Depending on flow
viscosity, flow rate,
there can be slightly
different sound speed
showing on different
channels. This could
be normal due to
different signal path.
•
Under ideal conditions, sound
speed spread of 30 ft/s (9 m/s)
or more between the sound
speed measurement of the
channels can be an indication
of a problem with the pipe
installation or any other
different local pipe condition.
SNR Up
Signal to noise ratio
of the upstream
transducer
>5
<2
SNR value between 2 and 5 shall
provide valid measurements but
can be an indication of a problem
with the pipe installation or any
other different local pipe condition.
Verify the clamping fixture
alignment, transducer spacing,
transducers, couplant, all the other
connections.
SNR Down
Signal to noise ratio
of the downstream
transducer
>5
<2
SNR value between 2 and 5 shall
provide valid measurements but
can be an indication of a problem
with the pipe installation or any
other different local pipe condition.
Verify the clamping fixture
alignment, transducer spacing,
transducers, couplant, all the other
connections.
Summary of Contents for Baker Hughes PanaFlow LC XMT1000
Page 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii PanaFlow LC User s Manual ...
Page 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Page 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Page 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Page 189: ......