PanaFlow™ LC User’s Manual
141
Appendix H. Foundation Fieldbus Communication
H.2.4 FISCO [Fieldbus Intrinsically Safe Concept] (cont.)
Note:
The XMT1000 FISCO control drawing is BH
GE drawing #752-584
. Please consult the factory for a copy
of the drawing.
IMPORTANT:
The FISCO cover on the XMT1000 terminal PCB is required to provide a barrier between IS and
non-IS connections. This cover must be installed if the Fieldbus application is FISCO.
The FISCO cover should come installed from the factory, as shown in
below. The Fieldbus cables
should be installed through the XMT1000 upper right port for direct entry into the FISCO zone on the
terminal PCB.
Figure 128: Installed FISCO Cover and Mounting Screw
Attention!
The FISCO cover must be installed to comply with FISCO guidelines.
Cable Port
Mounting Screw
(p/n 412-2884)
Cover (p/n 421-4378)