Appendix H. Foundation Fieldbus Communication
148
PanaFlow™ LC User’s Manual
H.4.3 NAMUR NE107 (cont.)
The
NAMUR NE107
Errors
and their
Default Categories
in the XMT1000
Resource Block
are listed in
below.
Table 36: NAMUR NE107 Errors and XMT1000 Default Categories
Error
Sub-Error Description
Default Category
CRC Configuration Error
Persistent Parameter CRC Fault
Failed
Software Failure
Stack Overflow failure
Failed
Sequence or Windowed Watch Dog
failure
Failed
Software fault
Failed
Device Initialization Failure
Initialization failed
Failed
Hardware Failure
ADC Bit Test Fault
Failed
VGA test Fault
Failed
Clock Frequency Fault
Failed
CPU test Fault
Failed
Invariable Flash Memory Fault
Failed
Variable Memory Fault
Failed
FPGA Configuration CRC fault
Failed
Temperature Test Fault
Failed
Driver Failure
Failed
Watch Dog failure
Failed
DSP Hardware Errors
Failed
Default ISR (DSP Exception)
Failed
DSP Exception
Failed
Modbus Communication Loss
No Modbus communication
Failed
Sensor Value out of Range
VelocityWarning
Offspec
Device not calibrated
NotCalibrated
Offspec
Sensor Measurement Error
SingleChAccuracy
Offspec
MultiChAccuracy
Offspec
ActiveTw
Offspec
CycleSkip
Offspec
Amplitude
Offspec
SignalQuality
Offspec
VelocityRange
Offspec
SoundSpeed
Offspec
SNR
Offspec
Device in Configuration Mode
In Configuration Mode - Indication
Check
Unsupported Parameter
Warning
Unsupported Parameter - Warning
Check
Summary of Contents for Baker Hughes PanaFlow LC XMT1000
Page 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii PanaFlow LC User s Manual ...
Page 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Page 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Page 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Page 189: ......