Chapter 2. Installation
16
PanaFlow™ LC User’s Manual
Figure 19: Transducer Mounting, Step 1
2.
Apply a thin bead of couplant to one of the transducers. A bead approximately the size of a
toothpaste bead should be placed down the center of the transducer face.
Figure 20: Transducer Mounting, Step 2
BHGE supplies an ultrasonic couplant for your clamp-on installation. The purpose of the couplant is to
provide reliable transmission of ultrasound between two adjacent solid surfaces. Couplants perform this
task by excluding air from the space between the adjacent surfaces. Accordingly, the clamp-on transducers
should be pressed tightly against the pipe, using hand pressure on the set screw to squeeze the couplant to as
thin a film as practical for the given pipe surface.
BHGE provides couplants for both permanent and temporary use as well as for high-temperature and
low-temperature applications. For long-term installations, make sure the couplant does not dry out or
extrude. Standard couplants supplied from BHGE for the
PanaFlow™ LC
are listed in Table 1 below.
Table 1: Standard BHGE Couplants
Part No.
Type
Temp. Range
Use
CPL-1
Standard
-40 to +240°C
Semi-Permanent
CPL-8
Solid Sheet
-40 to +260°C
Permanent
Summary of Contents for Baker Hughes PanaFlow LC XMT1000
Page 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii PanaFlow LC User s Manual ...
Page 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Page 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Page 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Page 189: ......