Revision 4
44
12/15/2018
a)
Retrieve Chromatogram
– presents a standard Windows
Open
dialog box
to
download
saved files from the chromatograph (*.bin).
b)
Upload Chromatogram
- presents a standard Windows
Open
dialog box
to send
files to
the chromatograph.
c)
Analyze Chromatogram -
determines the concentration of the various gases and
calculates the relevant parameters using the present configuration.
Retrieving and uploading of chromatograms is normally done as part of the
troubleshooting of the system. The user can change the parameters of the
component table and determine what the results are with the revised
component table values. These calculations are not archived or available
for reports.
d)
Show Analysis Markers -
displays markers on the chromatogram that indicate the start
and stop points for the integration of each peak, as well as the retention time for each
peak
.
e)
View Chromatogram –
present a standard Windows dialog box to select a chromatogram
file stored on the computer for retrieval and display on the tab.
f)
Factory Chromatogram -
presents a standard Windows
Open
dialog box to download a
factory calibration chromatogram.
•
Chart Functions – The icon at the extreme right of the ribbon bar is used to access a
variety of functions directly related to the chromatogram being collected. The selection
of the desired function can be performed by pressing the up or down arrows to the left
of the icon.
a) Pause Poll
- This stops the chromatogram from recording.
b) Delete Trace
- Removes the present active chromatogram(s).
c) Clean Chart
- Removes all chromatograms
d) Toggle a Trace
- Used to select the active trace if more than one trace is displayed
4.5.1.2 Right Clicking on the Plot
The
Chart
command and the
Archive Markers
command are presented if the right mouse
button is pressed when the mouse is in the Chromatogram grid.
•
Chart
Command (Figure 4-24).
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...