To Validate the System:
a) Place the analyzer into Reference mode via the
Stream Sequencer
selection of the
Configure G.C.
tab (Figure 7-1).
Figure 7-1: Stream Sequencer Selection of the Configure G.C. Tab
b) Set the top entry in
the Sequence of Runs
field to Ref 1 (or Ref 2 as required). If there
are additional entries in the table, delete them by moving the highlight to it (the active
entry will be highlighted in blue) and press the
Delete Sequence
button.
c) Transmit the changes to the analyzer by pressing the
Write to Analyzer
icon at the top of
the window. When the information has been transmitted, made the analyzer will
complete the run it is performing and then switch the appropriate solenoid to flow
calibration gas to the analyzer.
d) The analyzer will then inject the calibration gas and begin to analyze it. Once the run is
complete the results for the analysis can be viewed via the
Analysis Results
tab.
The analyzer will continue to analyze the calibration gas until the user
changes the Stream Sequencer back to the process stream(s).
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...