10.2.3 Instrumental Issues
Problem
Cause
Corrective action
No flow when Cal. Is
initiated.
Solenoid does not
energize
Check that there is 24VDC at
solenoid. Check that solenoid is wired
to connector P5 terminals 1 and 2. If
the solenoid still does not energize,
replace solenoid.
Analyzer won’t turn on.
No power
Check power termination connectors.
Check display power from
motherboard connector
P15(motherboard)-red, black going to
P10(display) –red, black
Communication Error is
displayed on the personal
computer
Communication setup is
incorrect.
Select the Connect To Analyzer icon
in the ACCUCHROME software.
Select the correct “Log in Navigation”
icon and follow the on screen
prompts.
Erratic flow meter float
Flow meter tube is
contaminated.
Clean flow meter tube with isopropyl
alcohol and dry with clean, dry
instrument air.