Revision 4
94
12/15/2018
There are two aspects to this page:
The
Enable Serial Ports
,
Expansion Serial Ports
and
Ethernet (TCP
) fields on the left are
used to indicate the port through which the data is to be exported. Each item in these three
regions can be toggled to either
Modbus 1
or
None
. Different Modbus lists can be assigned
to different communication ports.
The Enable Ports
check box should be selected if data is to be transmitted via a serial port.
The
Termination On
check box should be selected if a termination resistor is required for the
RS485 serial port.
The right side of the page is used to configure the port(s) that are used to transmit data. The
format of all ports except Ethernet Configuration is the same (Figure 8-5). The
communication parameters should be entered.
Figure 8-5: Setting Port Configuration
If the Ethernet Configuration is selected, the
Configuration
dialog box is as shown in Figure
8-6. The communication parameters should be entered.
Figure 8-6: Setting Ethernet Configuration
When the parameters have been selected, press the Write to Analyzer icon
Enter the address, port and pool time.
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...