Revision 4
3
12/15/2018
2.2 Chromatograph Oven
2.2.1 Injection/Switching Valves
The ACCUCHROME uses Valco Model DV22 injection/switching valves for directing the flow
of carrier gas and sample gas within the chromatograph oven. See Section 15 of this manual
for more information about the operation and maintenance of the valves.
2.2.2 Chromatograph Columns
The ACCUCHROME uses micro-packed columns for the separation of the natural gas into its
constituent components. The stationary phase is either porous polymer or liquid coated
diatomaceous earth and is packed into a 1/16” O.D. tube. The length of the column will
depend on the specific separation requirements.
The micro-packed columns are manufactured by Galvanic Applied Sciences. The
chromatograph valve and column are shown in Figure 2-2.
Figure 2-2: Chromatograph Valve and Columns
Column 2
Column 1
Sample Loop
Helium enters to
actuate valve