Revision 4
19
12/15/2018
Figure 3-4: Gas Tube-in Ports and Vent, Class 1, Div. 2 System
e) There are 2 vents from the oven on the ACCUCHROME. One vent , the “Sample Vent”
is used for the gas that flows through the chromatograph injection valve’s sample loop.
This line can be vented to atmosphere or to a low pressure flare header. The second
vent, the “Detector Vent”, is used for the effluent from the chromatograph column(s).
This must be vented to atmosphere.
f) Connect Signal Cables(s) to the analyzer. The analysis results are available on Modbus
Serial, Modbus TCP/IP, or by 4-20 mA signals.
See Section 12 for detailed wiring instructions.
Installation of the conduit, wiring and disconnect devices must comply with
all applicable national, local and user electrical codes.