Revision 4
77
12/15/2018
Section 6
Using the Instrument
6.1 Introduction
The ACCUCHROME is designed to monitor the composition of a gas stream on a
programmed basis. The system can analyze gas samples, a reference gas (gas of known
composition to verify acceptable system operation), and calibration gas (which is used to
calculate/recalculate response factors) as desired.
This section is provided to assist the user in the typical operation of the chromatograph and
should be used in conjunction with Chapters 4 and 5, which describe the general format of
the operating system.
It is assumed in this chapter that the system has been installed,
appropriate gas lines have been fitted and all connections to external
devices (e.g. alarms) have been made.
There are two modes of operation of the application software:
a)
Vie
w Mode - used to collect/view/Reporting data and viewing archived data (Section 6.2).
b)
Edit
Mode - used to set a broad range of parameters that define the data processing of
the data, define the system configuration (e.g. set alarms) and establish a schedule of
analyses (Section 6.3).
6.2 View Mode
Analytical data is collected via the
View
mode . In this mode, Reference streams, Calibration
streams and Sample streams are analyzed on a periodic basis as selected in Edit mode (data
can also be viewed while in the
Edit
mode). The analysis of a calibration stream is to adjust
the response factors used to determine the concentration of the various components of the
sample. Alarms will be activated upon various events and solenoids will be activated during
a separation as defined in Edit mode.
6.2.1 Viewing the Current Chromatogram
The chromatogram is presented on the monitor as it is being collected via the
Chromatogram
tab (Figure 6-1). The chromatogram presentation can be changed via the
Auto-Scale
icon
which sets the largest peak to 75% of full scale or the
Edit Chromatogram
icon which
presents a dialog box to define the X and Y minima and maxima as well as the X interval.
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...