Revision 4
57
12/15/2018
5.3 The Component Table
The
Component Table
consists of three regions:
•
A ribbon which contains controls for the table of chromatographic information and is used to
enter a number of values used in calculations. The nature of the ribbon is dependent on the
selection of the Method. If the ISO method is used, see Section 5.3.1.1 if the GPA or Liquid
method is used, see Section 5.3.1.2. .
•
A table for chromatographic parameters (Section 5.4)
•
A table that displays the calculated physical properties for the calibration gas. (Section 5.5)
5.3.1 The Ribbon
The
Ribbon
(Figure 5-2) is used to add/remove a line from tables, send a configuration to the
chromatograph and enter values required for calculations. Some typical values for all
components are found in the table in Section 14. The format of the ribbon is dependent on
the
Method
that is employed. If the GPA method is used, see Section 5.3.1.1, if the ISO
method is used, see Section 5.3.1.2.
Figure 5-2: The Ribbon (ISO Method)
5.3.1.1 The Ribbon – ISO Method
Add Component
– used to add a line to various tables as described below. The line is
added directly below the line indicated by the ►.
Delete Component
– is used to remove the line in the table indicated by the ►.
Write to Analyzer
– sends the present configuration to the analyzer.
The
Use Helium Fixed Factor
check box should be checked if it is desired to correct the
analytical results for the He concentration in the gas. The Helium concentration is to be
entered in the field below the
Use Unnormalized Concentration
field.
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...