Revision 4
83
12/15/2018
Section 7
ACCUCHROME Validation
7.1 Overview
The ACCUCHROME should be validated on a monthly basis or as per the measurement
requirements of the user.
The validation process involves analyzing a known gas mixture and comparing the analyzer
results to the certified value of the known mixture. This known mixture is referred to as the
calibration gas. The validation process is also used to assess the status of the
chromatograph and may be useful in indicating the presence of e potential problems such as
worn chromatograph valves or contaminated columns.
The calibration gas should be a mixture of all of the gasses the chromatograph is to measure
in proportions similar to those in the actual sample. The composition of a typical calibration
blend is shown in Table 7-1 for a C6+ measurement of natural gas. The calibration gas
should be obtained with a certificate of composition.
Table 7-1: Composition of a Typical Natural Gas Calibration Blend
Component
Mole %
Methane
Balance (89.67 %)
Ethane
5 %
Propane
1 %
n-Butane
0.3 %
iso-Butane
0.3 %
n-Pentane
0.1 %
iso-Pentane
0.1 %
n-Hexane
0.03 %
Nitrogen
1 %
Carbon Dioxide
2.5 %
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...